High Precision Analysis of Battery Materials

New uses for batteries, including providing power for everything from automobiles to medical devices to consumer products, are leading to rapid growth in their use. The growth of alternative energy sources also requires batteries to store excess energy and release it when needed.

New battery materials, technologies and types are constantly being developed as a result of this varied and rapid demand.

Some commonly-used elements in these new batteries include zinc (Zn), phosphorus (P), nickel (Ni), manganese (Mn) and lithium (Li), among others. Depending on the battery, the percentage of these components can range from up to 30% to just a single digit percent.

Accurately knowing the ratios of the major components is an important factor in order to obtain optimum performance because deviations (and the presence of impurities) can negatively affect the performance of a battery.

This means major battery components must be measured with precision and accuracy while simultaneously having their impurity levels measured. 

ICP-OES has a high tolerance to matrices but is still able to accurately measure impurities down to the μg/L level with high precision, making it the ideal analytical technique to meet these needs.

HP-ICP-OES, or high precision ICP-OES, provides unparalleled accuracy (RSDs of 0.1% or less) and is only available on fully simultaneous ICP-OES systems like PerkinElmer’s Avio® 550 Max.

This method measures the internal standard at the same time as the analyte for fully simultaneous measurements that compensate for all sources of noise.

This article will demonstrate the ability of the Avio 550 Mac ICP OES for measuring elements commonly found in batteries with both accuracy and high precision while simultaneously measuring common impurities.

Experimental

Materials and Sample Preparation

Table 1 shows the concentrations at which standards containing Zn, Ni, Li, P and Mn were prepared in 2% nitric acid (v/v) to mimic digestion of battery materials.

Common impurities were also added to solutions with the concentrations listed above. A 100x dilution of starting materials used in batteries was simulated by the concentrations shown in Table 1.

Table 1. Elements and Concentrations. Source: PerkinElmer, Inc.

Major Components
Element Concentrations (mg/L)
Mn, P 500, 1000, 2000
Li 125, 250, 500
Ni, Zn 50, 100, 200
Impurities
Element Concentrations (μg/L)
Cu, Cd, Hg, Mo, Pb 50, 100, 200

 

Instrumental Parameters

Analysis was completed using the parameters shown in Table 2, using the Avio 550 Max ICP-OES. Table 3 shows the timings, plasma views and wavelengths. Internal standards are read with the same integration time as the analytes for optimum precision with HP-ICP-OES.

More readings were able to be acquired by increasing precision through the use of a 10-second read time. Linear bracketing calibration curves were measured against the majors, and linear-through-zero regressions were used to measure impurities; all calibration standards used for measurement were prepared in 2% HNO3.

Table 2. Instrumental Parameters. Source: PerkinElmer, Inc.

Parameter Value
Nebulizer SeaSpray™
Spray Chamber Baffled Glass Cyclonic
Sample Uptake Rate 1.0 mL/min
RF Power 1500 W
Injector 2.0 mm id Alumina
Nebulizer Gas Flow 0.65 L/min
Auxiliary Gas Flow 0.2 L/min
Plasma Gas Flow 10 L/min
Torch Position -3
Replicates 3

 

Table 3. Method Parameters. Source: PerkinElmer, Inc.

Category Analyte Plasma
View
Internal
Standard
Plasma View
Height
Integration
Time (s)
Read
Time (s)
Majors Li 610.362 Radial Sc 361.383 20 0.01 10
Ni 232.003 Axial Sc 361.383 15 0.01 10
Zn 334.501 Axial Sc 361.383 15 0.01 10
Mn 347.413 Radial Sc 361.383 20 0.01 10
P 213.623 Radial Sc 361.383 15 0.01 10
Impurities Cd 214.440 Axial Sc 361.383 15 0.05 1
Cu 327.393 Axial Sc 361.383 15 0.05 1
Hg 194.168 Axial Sc 361.383 15 0.05 1
Mo 202.031 Axial Sc 361.383 15 0.05 1
Pb 220.353 Axial Sc 361.383 15 0.05 1

 

Results and Discussion

The middle standard was measured ten times against the constructed calibration curves. The RSDs, or relative standard deviations, for all the major elements of the samples are shown in Figure 1; they were all less than or equal to 0.2%.

Precision (%RSD) for major elements during 10 analyses of the middle calibration standard (Mn, P = 1000 mg/L; Li = 500 ppm; Ni, Zn = 100 ppm).???????

Figure 1. Precision (%RSD) for major elements during 10 analyses of the middle calibration standard (Mn, P = 1000 mg/L; Li = 500 ppm; Ni, Zn = 100 ppm). Image Credit: PerkinElmer, Inc.

Most of the results were <0.1%. This level of precision was enabled by the analyte being read at the exact same integration time (0.01 sec) as the internal standard.

The internal standard corrected for flicker and shot noise through having true simultaneous measurements, which resulted in exceptionally precise measurements.

Recovery of major elements over 10 analyses (Mn, P = 1000 mg/L; Li = 500 ppm; Ni, Zn = 100 ppm).

Figure 2. Recovery of major elements over 10 analyses (Mn, P = 1000 mg/L; Li = 500 ppm; Ni, Zn = 100 ppm). Image Credit: PerkinElmer, Inc.

Recovery of minor elements (0.1 mg/L) over 10 analyses.

Figure 3. Recovery of minor elements (0.1 mg/L) over 10 analyses. Image Credit: PerkinElmer, Inc.

Concentrations of the majors and impurities were both measured during the same analysis. The impurities are shown in Figure 3, and the recoveries for the majors are shown in Figure 2. The impurities recovered within 6% of their expected values, and the major elements recovered within 2%.

Table 4. Precision (% RSD) over 10 Analyses of Major and Impurity Elements. Source: PerkinElmer, Inc.

Major Components
Element RSD
Mn 0.12%
P 0.10%
Li 0.19%
Ni 0.07%
Zn 0.05%
Impurities
Element RSD
Cu 1.3%
Cd 1.2%
Hg 1.8%
Mo 1.3%
Pb 0.84%

 

The precision of both the major and impurity elements over the ten analyses evidence the stability of the methodology. Table 4 shows that the RSDs are less than 2% for the impurity elements and less than 0.2% for the major elements.

The use of HP-ICP-OES is the reason for the reproducibility of the major elements. 'Normal' and HP-ICP-OES analyses can both be implemented within the same method without the sacrifice of data quality, as demonstrated by the precision for the impurities.

Conclusion

This article shows how HP-ICP-OES can be provided using the Avio 550 Max ICP-OES through its ability to fully measure internal standards and analytes simultaneously and with high precision.

This results in high precisions/low RSDs (usually <0.10%) through its ability to provide real-time corrections of variations in the analytical signal.

Both impurities and matrix components can be measured together in a single method because HP-ICP-OES can be implemented in the same method along with analytes measured conventionally.

This is demonstrated in the above analysis of impurities and matrix components in common battery materials. HP-ICP-OES allows for exceptional accuracy and precision in the measurement of matrix materials when combined with bracketing calibrations.

Reference

  1. "HP-ICP-OES: Using the Avio 550/560 Max to Achieve the Highest Possible Precisions", Technical Note, PerkinElmer, 2020.

Consumables Used

Table 5. Source: PerkinElmer, Inc.

Component Part Number
Sample Uptake Tubing:
Black/Black (0.76 mm id) PVC, Flared
N0777043
Drain Tubing:
Gray/Gray (1.30 mm id), Santoprene
N0777444
Lithium Standard, 10,000 μg/mL N9304323 (125 mL)
N9304324 (500 mL)
Manganese Standard, 10,000 μg/mL N9304115 (125 mL)
Nickel Standard, 10,000 μg/mL N9304117 (125 mL)
N9304116 (500 mL)
Phosphorus Standard, 10,000 μg/mL N9304119 (125 mL)
N9304118 (500 mL)
Zinc Standard, 10,000 μg/mL N9304129 (125 mL)
Cadmium Standard, 1000 μg/mL N9300176 (125 mL)
N9300107 (500 mL)

 

Table 6. Source: PerkinElmer, Inc.

Component Part Number
Copper Standard, 1000 μg/mL N9300183 (125 mL)
N9300114 (500 mL)
Lead Standard, 1000 μg/mL N9300175 (125 mL)
N9300128 (500 mL)
Mercury Standard, 1000 μg/mL N9300174 (125 mL)
N9300133 (500 mL)
Molybdenum Standard, 1000 μg/mL N9303784 (125 mL)
N9300134 (500 mL)
Scandium Standard, 1000 μg/mL N9303798 (125 mL)
N9300148 (500 mL)
Autosampler Tubes, case of 500 B0193233 (15 mL)
B0193234 (50 mL)

 

This information has been sourced, reviewed and adapted from materials provided by PerkinElmer, Inc.

For more information on this source, please visit PerkinElmer, Inc.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    PerkinElmer, Inc.. (2024, February 29). High Precision Analysis of Battery Materials. AZoCleantech. Retrieved on October 30, 2024 from https://www.azocleantech.com/article.aspx?ArticleID=1212.

  • MLA

    PerkinElmer, Inc.. "High Precision Analysis of Battery Materials". AZoCleantech. 30 October 2024. <https://www.azocleantech.com/article.aspx?ArticleID=1212>.

  • Chicago

    PerkinElmer, Inc.. "High Precision Analysis of Battery Materials". AZoCleantech. https://www.azocleantech.com/article.aspx?ArticleID=1212. (accessed October 30, 2024).

  • Harvard

    PerkinElmer, Inc.. 2024. High Precision Analysis of Battery Materials. AZoCleantech, viewed 30 October 2024, https://www.azocleantech.com/article.aspx?ArticleID=1212.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.